In brief
- A prospective European multicentre study tested a standardised cleaning protocol for transbronchial needle aspiration (TBNA) needles reused between targets during EBUS and EUS-B.
- Among malignant cases, needle rinse samples taken after cleaning were negative for malignant cells in 92.2% and positive or suspicious in 7.8%.
- Cleaning took a mean of 148.8 seconds per target, and no complications were observed.
- The needle looked clean in 89.3% of cases, but visual assessment did not correlate with the cytology of the rinse.
- No clinical or procedural variable predicted which cleanings would fail.
Study at a glance
- Design: prospective, multicentre, European
- Population: 162 of 164 enrolled patients undergoing EBUS and/or EUS-B for diagnosis or staging of lung cancer or metastatic disease
- Targets sampled: 270; malignancy confirmed in 76.3%
- Intervention: standardised TBNA needle-cleaning protocol between targets
- Primary endpoint: no malignant cells in the post-cleaning needle rinse
- Main result: rinse negative in 92.2% of malignant cases, positive or suspicious in 7.8%
- Safety: no complications observed
- Not reported: in the abstract, the number of centres and the steps of the protocol
Needle reuse between N2 stations as the open question
EBUS and EUS-B are standard tools for mediastinal staging, and the 9th edition of the TNM classification gives weight to station-by-station nodal assessment. In principle, when the same needle samples more than one N2 station, malignant cells carried over from a positive node could make a negative node read as positive and upstage the patient. The authors describe the best sampling strategy, and needle reuse in particular, as unresolved.
A clean rinse in 92.2% of malignant cases
Of 164 patients enrolled, 162 entered the final analysis, contributing 270 sampled targets. The needle was cleaned with the standardised protocol, and post-cleaning rinse samples were assessed cytologically. In malignant cases this rinse was free of malignant cells 92.2% of the time, while 7.8% of rinses came back positive or suspicious.
Cleaning took 148.8 ± 40.9 seconds per target on average. No complications were observed.
A needle that looks clean is not a reliable guide
The needle was visually assessed as clean in 89.3% of cases, and that assessment did not correlate with the rinse cytology. No clinical or procedural variable was associated with a failed cleaning either, so the study offers no way to pick out in advance the cases in which carry-over persists.
“However, residual contamination persists in a minority of cases, indicating that needle reuse may carry a small risk of cross-contamination between targets.”Bugalho et al., Pulmonology (2026)
An alternative to routine needle exchange, with a residual risk
The authors conclude that the protocol is safe and effective in reducing contamination, and that it may be a pragmatic alternative to exchanging the needle as a matter of routine. The rinse endpoint measures cells left on the needle after cleaning; it does not measure how often those cells reached the next node and altered the staging result.
The figures set roughly two and a half minutes of cleaning per target against a positive or suspicious rinse in about one malignant case in thirteen.